2021 EXTRAKTLAB e 140

2021 EXTRAKTLAB e 140

Contact us for price

Location:Poland

Description

Manufacturer: EXTRAKTLAB

Model: e 140

Condition: Refurbished (2021)

Processes up to 8 kg of input material per hour depending on process conditions.

Extraction vessels: 2 x 20 L

Collection vessels: 3 x 2.5 L

Recycler vessels: 1 x 20 L

CO2 Pump Flow Rates Range (LPM): 1.5 to 2 LPM

Max Operating Pressure (psi): 5000 psi, 344.74 bar

Maximum Process Temperature: 90°C

Skid Operating Temperature (C°): 5 to 40°C (15 to 30°C optimal)

CO2 Solvent Inlet Lines: 1

Eluent Inlet Lines: up to 2

User Interface: Touch Screen HMI panel with barcode scanner, maintenance tracking, recipe storage (parameter). 21 CFR Part 11 compliant software.

Biomass Loading and Unloading: 10 micron flexible no channel bags

Operating Humidity: 20 to 80%

Compliance – UL 508A electrical panel Electrical – 50 Hz 208V-240V, 3 phase,

Footprint:- 24ft. Dimensions – 99″ x 34″ x 72″ (LxWxH)

Construction: 316 stainless steel, Ferrule compression fittingsthroughout

Sound Pressure – 50 dB

Wetted Materials – FDA approved structural stainless steel.

Chemical Compatibility: CO2, Organic Solvents, Acids and Bases, plant extracts

Flow Control: Manual, Automatic

Flowe Meter: PID Controlled Algorithm, rules based on process parameters

CE and food grade GMP certified

Independently PE Stamped according to ASME DIV II VII-2PED compliant, CE marked, notified body, Lloyd's Register, CRN number available

Safety Specifications:

Rupture Discs: Zook burst discs, calibrated, ASME VIII Div IIcompliant, Max Pressure is 5300 psi, Certificate Attached to Disc

Temperature Switch: Preheater, Post heater MAX TEMP 140°C,Certificate Available Collector, Extractors MAX TEMP 100°C, Certificate Available

E-stop Switch: Switch disengages main contactor, not dependenton PLC

CO2 Gas Detector: Built in alarm, turns off pump and closes valves at9000

Interlocks: Marked with safety warnings

Pressure Transducers: Post pump, Extractor, Collector 1.3; Turns off pumpwhen overpressure (>5100 psi) condition is detected

Isolation Valves: Manual, bypass valves.

Technical documentation: available.



Find more here - https://www.ucymachines.com/


Contact us at - sales@ucymachines.com

Specifications

ManufacturerEXTRAKTLAB
Modele 140
Year2021
ConditionUsed
Stock Number756732